Replacement - Kodak X-Omat 2000 Service Manual

Hide thumbs Also See for X-Omat 2000:
Table of Contents

Advertisement

SERVICE MANUAL
[3] Install:
• DEVELOPER RACK
• DETECTOR CROSSOVER AY
[4] Loosen the 2 SCREWS on the DETECTOR SWITCHES.
[5] Move the 2 DETECTOR SWITCHES up to the top position.
[6] Set the distance between the MAGNET and the DETECTOR SWITCH on the drive side to 1.6 mm (0.06 in.):
(a) Loosen the SCREW.
(b) Move the ROCKER ARM.
(c) Tighten the SCREW.
Warning
Dangerous Voltage
[7] Energize the PROCESSOR.
[8] Lift the top DETECTOR ROLLER.
[9] Insert a 2.54 cm (1 in.) piece of film into the drive side of the DETECTOR ROLLER.
[10] Slowly move the DETECTOR SWITCH on the drive side down until the REPLENISHMENT PUMP actuates.
[11] Tighten the 2 SCREWS on the DETECTOR SWITCH.
[12] Check that the REPLENISHMENT PUMP stops in 3 seconds after you remove the film.
Note
The feed signal will emit a beep.
[13] Repeat Steps 8-12 for the DETECTOR SWITCH on the non-drive side.
[14] Install the TOP COVER.
[15] Check that the REPLENISHMENT PUMP does not operate unless you feed film.

Replacement

Warning
Dangerous Voltage
[1] Disconnect the PROCESSOR from the main power.
6
2 LOCK WASHERS
2 SCREWS
DETECTOR
SWITCH
2 WASHERS
H172_1037BCA
H172_1037BA
12JAN99 – 7C8772

Hide quick links: