5
Quick Start Guide
Data Acquisition in Method and Run Control View

Running the Method for a Single Injection

This section shows how to run a single injection of the isocratic checkout
sample using the conditions entered in the previous section.
Analyses on the ChemStation can be run in two modes:
• Run Method — single injections, for example, in interactive method
• Run Sequence — automated series of injections from multiple vials,
1 Click the Select Run Method Task icon
2 If required method conditions are not currently loaded, select Method >
If changes have been made to a loaded method and not yet saved, this is indicated by a
N O T E
yellow asterisk on the method status icon. The injection can be made without first saving
those parameters changes. The ChemStation always stores a copy of the acquisition
parameters in the data file as ACQ.TXT to ensure the preservation of the original method
parameters.
3 Place the sample vial in position 1 — this is the front position of the 10
4 Select Run Control > Sample Info, and enter the required sample
5 If the system is already equilibrated and the baseline is stable, click Run
94
development, using the parameter settings current at the time;
possibly with multiple methods. For further details refer to the
ChemStation manuals.
Load Method or the icon
x 2 ml vial positions on the right hand side of the sample tray.
parameters.
For example:
Subdirectory (optional)
Name Pattern
Vial/Location
Sample name
Comment
Method in the Sample Info page to start the injection. Alternatively, click
OK, and when ready, select Run Control > Run Method. The item will be
added to the Run Queue and then automatically started from there.
.
under the menu bar to load them.
1260 Infinity Binary LC - System User Guide