Agilent Technologies Agilent 1260 Infinity System User's Manual page 20

Binary lc
Table of Contents

Advertisement

1
The Agilent 1260 Infinity Binary LC - Product Description
System Components
The chromatographic signals are extracted from the diode- array data
within the firmware of the module. Up to eight individual signals can be
defined. Each of them comprises a signal wavelength, a diode- bunching
bandwidth and - if required - a reference wavelength and bandwidth.
Signals can be output at 80 Hz (80 data points/second) for accurate
recording of the fastest (narrowest) chromatographic peaks. At the same
time, the module can also output full- range spectra to the data system at
the same rate of 80 Hz.
For regulated laboratories, it is important that all the method parameters
are recorded. The 1260 Infinity DAD not only records the instrument
setpoints but also has RFID tags (radio- frequency identification tags)
incorporated into the lamp and flow cell cartridge so that the identity and
variables of these important components are also recorded by the system.
20
1260 Infinity Binary LC - System User Guide

Advertisement

Table of Contents
loading

Table of Contents