Agilent Technologies Agilent 1260 Infinity System User's Manual page 25

Binary lc
Table of Contents

Advertisement

Table 2
Type
Composition
precision
Composition
accuracy
Control
Local control
Analog output
Communications
Safety and
maintenance
GLP features
Housing
For use with flow rates below 500 µl/min or for use without damper and mixer a vacuum
N O T E
degasser is required.
All specification measurements are done with degassed solvents.
1260 Infinity Binary LC - System User Guide
The Agilent 1260 Infinity Binary LC - Product Description
Performance Specifications of the Agilent 1260 Infinity Binary Pump (G1312B)
Specification
< 0.15 % RSD or < 0.04 min SD whatever
is greater
± 0.35 % absolute, at 2 mL/min, at
10 MPa (100 bar)
Agilent control software (e.g.
ChemStation, EZChrom, OL, MassHunter)
Agilent Instant Pilot
For pressure monitoring, 1.33 mV/bar,
one output
Controller-area network (CAN), RS-232C,
APG Remote: ready, start, stop and
shut-down signals, LAN optional
Extensive support for troubleshooting and
maintenance is provided by the Instant
Pilot, Agilent Lab Advisor, and the
Chromatography Data System.
Safety-related features are leak detection,
safe leak handling, leak output signal for
shutdown of pumping system, and low
voltages in major maintenance areas.
Early maintenance feedback (EMF) for
continuous tracking of instrument usage
in terms of seal wear and volume of
pumped mobile phase with pre-defined
and user settable limits and feedback
messages. Electronic records of
maintenance and errors
All materials are recyclable
Specifications
Comments
at 0.2 and 1 mL/min; based
on retention time at constant
room temperature
(water/caffeine tracer)
Revision B.02.00 or above
1
25