Agilent Technologies Agilent 1260 Infinity System User's Manual page 78

Binary lc
Table of Contents

Advertisement

4
System Setup and Installation
Installing the Modules
Regular Priming
When
When the pumping system has been turned off for a certain time (for example, overnight) air will
rediffuse into the solvent channel between the vacuum degasser and the pump. If solvents
containing volatile components are left in the degasser without flow for a prolonged period, there
will be a slight loss of the volatile components.
Preparations
Switch the system on
The purge tool of the Lab Advisor can be used for automatically purging the pump.
N O T E
1 Open the purge valve of your pump by turning it counterclockwise and
2 Flush the vacuum degasser and all tubes with at least 10 mL of solvent.
3 Repeat step 1 and 2 for the other channel(s) of the pump.
4 Set the required composition and flow rate for your application and
5 Pump for approximately 10 minutes before starting your application.
78
set the flow rate to 5 mL/min.
close the purge valve.
1260 Infinity Binary LC - System User Guide