Integrating a Signal

1 Select the Integration Task Tool (see figure below). The Integrate icon
2 Click the Set Integration Events Table icon
3 Set Baseline Correction to Advanced for gradient runs.
4 Set Slope Sensitivity to 5.
5 Set the Peak Width value to the narrowest peak of interest,
6 Area Reject and Height Reject can be set to reject the smallest peaks.
7 Click the Integrate icon
8 Exit from the Integration Events table using the green tick icon
98
and the Set Integration Events Table icon are highlighted in the figure
shown below.
Higher numbers will integrate steeper peaks and ignore less steep
peaks.
approximately 0.02 in this case.
settings.
to open the table as shown.
to update the results using these new
1260 Infinity Binary LC - System User Guide
.