Purging the Pump

Purge the pump, if ...
• The pump has been primed for the first time.
• The pump is to be purged with fresh solvent before using the system, or
• The pump has been idle for a few hours or more (air may have diffused
• The solvent reservoirs are refilled, and the pump requires purging to fill
For details on the purging procedure, refer to
page 78.
1260 Infinity Binary LC - System User Guide
when the solvent is to be exchanged for another.
into the solvent lines and purging is recommended).
the system with fresh solvent. If different solvents are to be used,
ensure that the new solvent is miscible with the previous solvent and if
necessary use an intermediate step with a co- miscible solvent
(isopropanol is often a good choice, check with a solvent miscibility
table).
Quick Start Guide
Preparing the System
"Regular Priming"
on
5
87