Agilent Technologies Agilent 1260 Infinity System User's Manual page 123

Binary lc
Table of Contents

Advertisement

• Column Switching Valve is an active option only when a valve is fitted
1260 Infinity Binary LC - System User Guide
The temperature of each zone can be set from - 5 °C to 100 °C and the
user should check that the column is suitable for operation at that
temperature. (Agilent ZORBAX RRHT StableBond phases can be used at
the higher end of the range). The temperature is controlled to ± 0.15 °C
down to 10 °C below ambient although it should be noted that there
are very few applications operating below 12- 15 °C. Avoid using the
TCC at such low temperature that condensation of water from humid
air occurs as this will trigger the leak sensor.
between the column holders. Three types of valve are available:
2- position 6- port – used to switch between 2 columns
2- position 10- port – used for alternating column regeneration
8- position 9- port – used for multiple column selection in MDS
Setting Up a Method using Edit Entire Method
6
Appendix
123

Advertisement

Table of Contents
loading

Table of Contents