Autosampler tab
Figure 25
• Injection Mode
• Needle Wash – if selected above.
Vials should not have a septum i.e., they should be open to avoid transfer of carryover
N O T E
material on the septum.
1260 Infinity Binary LC - System User Guide
Setup Method screen – High Performance Autosampler tab
Injection volume sets the volume to be injected (example 1 µl),
Standard injection indicates that no external needle wash is done,
Injection with needle wash is used to reduce potential carryover. This is
the recommended option and is configured in the next entry.
Mode determines how the outside of the needle will be rinsed: either
actively in the Flush Port or by dipping into a specified Wash Vial.
Time in seconds the peristaltic pump connected to the flush port will
pump the rinse solvent. It then pumps for a further 15 s to clear the
flush port.
Location determines which vial or well plate will be used if Wash Vial
was selected.
Setting Up a Method using Edit Entire Method
6
Appendix
115