Agilent Technologies Agilent 1260 Infinity System User's Manual page 64

Binary lc
Table of Contents

Advertisement

3
Optimization of the Agilent 1260 Infinity Binary LC
How to Achieve Higher Sensitivity
Ultra-Low Dispersion
The Max- Light Cartridge ULD cell can be used with the G4212A DAD and
G4212B DAD . The cell is a requirement for the Ultra- Low Dispersion Kit
solution which currently exists as 1290 Infinity Ultra- Low Dispersion Kit
(5067- 5189). The cell should be part of the ultra- low dispersion solution.
High Dynamic Range
The Max- Light Cartridge HDR cell can be used with the G4212A DAD and
G4212B DAD . The cell is required as a part of the High Dynamic Range
(HDR) solution which will be introduced March/April 2013.
To protect the flow cell against overpressure (e.g. in systems with LC/MS) install Inline
N O T E
Pressure Relief Valve Kit (G4212-68001), see Agilent 1200 Infinity Series Diode Array
Detectors manual.
64
1260 Infinity Binary LC - System User Guide