Agilent Technologies Agilent 1260 Infinity System User's Manual page 90

Binary lc
Table of Contents

Advertisement

5
Quick Start Guide
Data Acquisition in Method and Run Control View
Table 10
Module
Pump
Autosampler
Column
Compartment
Detector
Checkout specifications:
• four peaks are visible in the chromatogram
• Intensity Test – pass
• Wavelength Calibration±1 nm
90
Method parameters for first separation run - low delay volume configuration
Parameter
Solvent A
Solvent B
Flow rate
Solvent
Composition
Stop Time
Injection
Column
Temperature
Signal
Setting
Water
Acetonitrile
0.75 mL/min
40 % A, 60 % B
1.2 min (1 min post time)
1 µL
Column: SB-C18, 4.6x 50 mm, 1.8 µm, 600 bar
(827975-902)
40 °C
Wavelength: 250 nm, Bandwidth: 4 nm
Reference Wavelength: 360 nm, Reference
Bandwidth: 100 nm
1260 Infinity Binary LC - System User Guide