Optimization of the Agilent 1260 Infinity Binary LC
How to Configure the Optimum Delay Volume
The standard delay volume configuration is also the configuration which
provides direct method transferability from the Agilent 1100 and 1200
Series LC system to the 1260 Infinity Binary LC or vice versa. Figure
Figure 16
analysis of paracetamol and related impurities. The method was
transferred from an Agilent 1200 Series LC system to an Agilent 1260
Infinity Binary LC with the chromatographic conditions (column, mobile
phase, pump settings, injection volume, column temperature, detector
settings) left unchanged. As can be seen, a seamless method transfer is
possible.
Figure 16
46
on page 46 shows two overlaid chromatograms of a method for
Comparison Agilent 1200 Series and Agilent 1260 Infinity LC system on one
Configuration
1260 Infinity Binary LC - System User Guide