Archon Connections
to Varian 3400 and 3600 GC's with Star Workstation
Remote I/O Cable Hookup Instructions
Archon to Tekmar LSC 2000/3000 to Varian 3400 and
3600 GC's
Archon Connections to Varian 3400 and 3600 GC's with Star Workstation
Remote I/O Cable Hookup Instructions:
Archon to Tekmar LSC 2000/3000 to Varian 3400 and 3600 GC's
When connecting to a Varian series 3400 or 3600 GC with Star Workstation,
connectors from both the Archon I/O cable and the Star Workstation cable
must access the J16 and J23 connectors on the GC motherboard. This is
possible using the J16 Splitter (p/n DY-700030-00) and the J23 Splitter
(p/n DY-700031-00) following the instructions below:
Turn off the power and unplug the power cord from the GC. Remove the
1
top panels to allow access to the GC motherboard.
Route the GC-end of the Archon I/O cable and the GC-end of the Star
2
Workstation cable through the back panel of the GC (see Star Workstation
installation diagram).
3
Connect the 5-pin female Molex connector (P23) from the Star Workstation
cable to the male connector on the J23 Splitter with 4 pins (the second
position lacks a pin).
4
Connect the 5-pin female Molex connector from the Archon I/O cable to the
5-pin male connector on the J23 Splitter.
Connect the J23 Splitter to the J23 connector on the GC motherboard (see
5
GC manual for motherboard layout schematic).
Connect the 4-pin female Molex connector from the Archon I/O cable to
6
either of the 4-pin male connectors on the J16 Splitter.
7
Connect the 4-pin female Molex connector (P16) from the Star Workstation
cable to either of the 4-pin male connectors on the J16 Splitter.
Agilent Technologies

Advertisement

loading

Summary of Contents for Agilent Technologies Archon

  • Page 1 Turn off the power and unplug the power cord from the GC. Remove the top panels to allow access to the GC motherboard. Route the GC-end of the Archon I/O cable and the GC-end of the Star Workstation cable through the back panel of the GC (see Star Workstation installation diagram).
  • Page 2 Complete the installation of the Archon I/O cable and of the Star Workstation cable according to the standard instructions for those cables. The following sequence occurs when the Archon and a Purge and Trap unit is NO TE configured with the Varian 3400 and 3600 GC with Star Workstation: The Archon begins its sample handling after receiving a Ready signal from the Purge and Trap unit.
  • Page 3 Archon Connections Remote I/O Cable Hookup Instructions...
  • Page 4 © Agilent Technologies, Inc. 2012 Printed in USA First edition, March 2012 Agilent Technologies, Inc.