Optimum Instrument Configuration for 3 and 4.6 mm i.d. Columns

Standard delay volume configuration for highest UV sensitivity and direct
method transferability
The relative column volumes for 3 mm and 4.6 mm inner diameter
columns are about two and five times larger respectively than for the
same length 2.1 mm i.d. columns, and the flow rates that are used are
also proportionally higher. Therefore, the standard delay volume of the
binary pump will not result in a significantly higher gradient delay.
Figure 15
1260 Infinity Binary LC - System User Guide
Optimization of the Agilent 1260 Infinity Binary LC
Standard delay volume configuration for 3 and 4.6 mm ID columns with
highest UV sensitivity
How to Configure the Optimum Delay Volume