The Agilent 1260 Infinity Binary LC - Product Description
Specifications
Performance Specifications
Table 2
Type
Hydraulic system
Setable flow range
Flow range
Flow precision
Flow accuracy
Pressure operating
range
Pressure pulsation
Compressibility
compensation
Recommended pH
range
Gradient formation
Delay volume
Composition range
24
Performance Specifications of the Agilent 1260 Infinity Binary Pump (G1312B)
Specification
Two dual piston in series pumps with
servo-controlled variable stroke drive,
power transmission by gears and ball
screws, floating pistons
Set points 0.001 – 5 mL/min, in
0.001 mL/min increments
0.05 – 5.0 mL/min
≤0.07 % RSD or ≤0.02 min SD, whatever is
greater
± 1 % or 10 µL/min, what ever is greater
Operating range 0 – 60 MPa (0 – 600 bar,
0 – 8700 psi) up to 5 mL/min
< 2 % amplitude (typically < 1.3 %), or <
0.3 MPa (3 bar), whatever is greater, at
1 mL/min isopropanol, at all pressures >
1 MPa (10 bar, 147 psi)
Low delay volume configuration:
< 5 % amplitude (typically < 2 %)
Pre-defined, based on mobile phase
compressibility
1.0 – 12.5, solvents with pH < 2.3 should
not contain acids which attack stainless
steel
High-pressure binary mixing
Standard delay volume configuration:
600 – 800 µL, (includes 400 µL mixer),
dependent on back pressure
Low delay volume configuration:
120 µL
settable range: 0 – 100 %
recommended range: 1 – 99 % or
5 µL/min per channel, whatever is greater
Comments
based on retention time at
constant room temperature
pumping degassed H
10 MPa (100 bar)
measured with water at
1 mL/min (water/caffeine
tracer)
1260 Infinity Binary LC - System User Guide
O at
2