How To Achieve Shorter Cycle Times - Agilent Technologies Agilent 1260 Infinity System User's Manual

Binary lc
Table of Contents

Advertisement

3
Optimization of the Agilent 1260 Infinity Binary LC

How to Achieve Shorter Cycle Times

How to Achieve Shorter Cycle Times

Shorter cycle times can be achieved by selecting a short column with good
selectivity. The column dimensions are also determined by the detection
system that is used. For UV detection, 3.0 mm inner diameter columns are
ideal, because here the highest linear velocities can be obtained. With
4.6 mm inner diameter columns, high linear velocities can also be reached,
but the maximum flow rate is limited to 5 mL/min.
The pump should be used in its standard delay volume configuration (see
Figure 13
diameter columns. For 2.1 mm inner diameter columns, the low delay
volume configuration should be used. In addition, when using 2.1 mm
inner diameter columns, the low dispersion kit should be installed to
provide lowest extra- column volume. For highest UV sensitivity, it is
recommended in addition to use the short mixer ( Low volume mixer (
200 µL) (5067- 1565)).
Chromatographic conditions strongly depend on the compounds that are
to be analyzed, but some rules of thumb can be used to achieve short run
times:
• The flow rates should be as high as possible, depending on the required
• Steep gradients should be used.
• High column temperatures are recommended to enable high flow rates
Alternating column regeneration
Even shorter cycle times can be achieved by using a column regeneration
valve in combination with a regeneration pump. Using this setup, the
regeneration of the column that is used previously takes place during the
analysis on the second column. This shortens cycle time significantly.
Using two columns, two pumps and one 2- position 10- port valve allows
switching between these columns for shortest cycle times from injection to
injection. Typically, columns of the same chemistry and the same batch
provide a retention time precision that allows data processing using the
same calibration table.
Please refer to the Valve Solution User Guide for a more detailed
description of the alternating column regeneration.
48
on page 43) for 4.6 mm inner diameter and 3.0 mm inner
resolution, back- pressure and the detection system used.
to be used, and to shorten run time even further. Zorbax SB columns
can be used up to 90 °C, at low pH.
1260 Infinity Binary LC - System User Guide

Advertisement

Table of Contents
loading

Table of Contents