Delay Volumes In The Agilent 1260 Infinity Binary Lc - Agilent Technologies Agilent 1260 Infinity System User's Manual

Binary lc
Table of Contents

Advertisement

3
Optimization of the Agilent 1260 Infinity Binary LC
How to Configure the Optimum Delay Volume

Delay Volumes in the Agilent 1260 Infinity Binary LC

Table 4
which contribute to system delay volume in the Agilent 1260 Infinity
Binary LC System .
Table 4
Components
Binary Pump
Binary Pump
Low volume mixer
Mixer
Autosampler
Low dispersion heat exchanger
Built-in heat exchanger
1
2
Table 5
System Configuration
Low delay volume configuration
Medium delay volume configuration
Standard delay volume configuration
Switching between configurations can be done in two ways:
• manually, by disconnecting and reconnecting capillaries
• automatically, using a 600 bar 2PS/6PT valve (optional)
42
on page 42 and
Delay volumes of the 1260 Infinity Binary LC modules
1
2
in low delay volume configuration with bypassed damper and mixer
in standard delay volume configuration
Delay volumes of 1260 Infinity Binary LC configurations
Table 5
on page 42show the component volumes
Delay Volume (µL)
120
600 – 800
200
400
270
1.6
3 and 6
Delay Volume (µL)
Pump: 120
Autosampler:
Pump: 320
Pump: 600 – 800
1260 Infinity Binary LC - System User Guide

Advertisement

Table of Contents
loading

Table of Contents