Controlling The Agilent 1260 Infinity Ii Sfc Control Module Through The Agilent Openlab Cds - Agilent Technologies InfinityLab LC Series User Manual

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

Controlling the Agilent 1260 Infinity II SFC Control Module through the Agilent OpenLAB CDS

Controlling the Agilent 1260 Infinity II SFC Control Module
through the Agilent OpenLAB CDS
Figure 23
Editable method parameters for the Agilent 1260 Infinity II SFC Control
Module are the BPR Pressure (system back-pressure) and the BPR Temperature
(temperature of the back-pressure regulator). Via timetable the BPR Pressure
can be changed within the run and BPR Pressure gradients can be performed.
In addition, you can set a Stoptime for the analysis, as for all other modules.
For Hybrid Systems (SFC and HPLC in one system) there is the possibility to
choose the mode how the SFC Control module should behave. This function
should only be used in Hybrid systems, in pure SFC Systems the SFC Mode
check box has always to be checked. When the SFC Mode check box is checked
the SFC Control Module performs under SFC conditions; the BPR Pressure and
BPR Temperature are applied and the Booster Pump provides compressed CO
(8 bar below system pressure).
Using the Agilent 1260 Infinity II SFC Control Module
Editing the SFC control module Method Parameters
5
2
71

Advertisement

Table of Contents
loading

This manual is also suitable for:

1260 infinity ii sfc system

Table of Contents