Parameterization Overview (Factory Settings) - ABB ProcessMaster FEM630 Operating Instructions Manual

Electromagnetic flowmeter
Hide thumbs Also See for ProcessMaster FEM630:
Table of Contents

Advertisement

FE M63 0 | E L EC TR OM AG N E T I C FLOWM E T E R | O I/F E M6 3 0-EN RE V. A
. . .15
Appendix

Parameterization overview (factory settings)

Parameter
Sensor Tag
Sensor Location Tag
Qv Max 1
Unit Volumeflow Qv
Unit Vol. Totalizer
Pulses per Unit
Pulse Width
Damping
Operating mode Digital output 41 / 42
Operating mode Digital output 51 / 52
Curr.Out 31/32
Curr.Out at Alarm
Current at flow rate> 103 % (I=20.5 mA)
Low flow cutoff
Empty pipe detection
Value range
Alphanumeric, maximum 20 characters.
Alphanumeric, maximum 20 characters.
Depending on the nominal diameter of the sensor.
l/s; l/min; l/h; ml/s; ml/min; m3/s; m3/min; m3/h; m3/d; hl/h; g/s; g/min; g/h;
kg/s; kg/min; kg/h; kg/d; t/min; t/h; t/d
m3; l; ml; hl; g; kg; t
1 to 10000
0.1 to 2000 ms
0.02 to 60 s
Off, Binary output, Pulse output, Frequency output
Off, Binary output, pulse output (follows digital output 41 / 42, 90 ° or 180 °
out of phase)
4-20mA FWD/REV, 4-20mA FWD, 4-12-20 mA
High Alarm 21 to 23 mA or Low Alarm 3.5 to 3.6 mA
Off (current output remains at 20.5 mA), High Alarm, Low Alarm ,
0 to 10 %
On / Off
137
Factory setting
None
None
Set to Q
DN in accordance with
max
Measuring range table on page
59.
g/min
Gallon (g)
1
100 ms
1
Digital output 41/42 as pulse
output for forward flow and
reverse flow
Digital output 51 / 52 as binary
output for output of the flow
direction.
4-20mA FWD/REV
High Alarm, 21.8 mA
Off
1 %
Off