FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | CI/FEP630/FEH630-EN REV. D
Parameterization overview (factory settings)
Parameter
Sensor Tag
Sensor Location Tag
Qv Max 1
Unit Volumeflow Qv
Unit Vol. Totalizer
Pulses per Unit
Pulse Width
Damping
Operating mode Digital output 41 / 42 Off, Binary output, Pulse output, Frequency output
Operating mode Digital output 51 / 52
Curr.Out 31/32
Curr.Out at Alarm
Current at flow rate> 103 % (I=20.5 mA) Off (current output remains at 20.5 mA), High Alarm, Low
Low flow cutoff
Empty pipe detection
Software history
In accordance with NAMUR recommendation NE53, ABB offers a transparent and traceable software history.
Device software package FEx630 (device firmware package)
Design
Issue date
00.04.00
2/3/2017
00.04.01
6/27/2017
00:05:00
1/12/2018
7/1/2000
2018
Change from one to two columns
Value range
Alphanumeric, maximum 20 characters.
Alphanumeric, maximum 20 characters.
Depending on the nominal diameter of the sensor.
l/s; l/min; l/h; ml/s; ml/min; m3/s; m3/min; m3/h; m3/d;
hl/h; g/s; g/min; g/h; kg/s; kg/min; kg/h; kg/d; t/min;
t/h; t/d
m3; l; ml; hl; g; kg; t
1 to 10000
0.1 to 2000 ms
0.02 to 60 s
Off, Binary output, pulse output (follows digital output
41 / 42, 90 ° or 180 ° out of phase)
4-20mA FWD/REV, 4-20mA FWD, 4-12-20 mA
High Alarm 21 to 23 mA or Low Alarm 3.5 to 3.6 mA
Alarm ,
0 to 10 %
On / Off
Type of change
Description
First publication
–
Bug fixing
Piston pumps filter
Bug fixing
Integrated Polish language
Bug fixing
PROFIBUS DP® and Modbus® integrated. New bootloader
Set to Q
DN in accordance with Measuring range
max
Digital output 41/42 as pulse output for forward flow
Digital output 51 / 52 as binary output for output of
3KXF002044U0100_00.04.00
3KXF002044U0100_00.04.01
3KXF002044U0100_00.05.00
3KXF002044U0100_01.07.00
55
Factory setting
None
None
table on page 54.
l/min
Liter (l)
1
100 ms
1
and reverse flow
the flow direction.
4-20mA FWD/REV
High Alarm, 21.8 mA
Off
1 %
Off
Ordering number
Need help?
Do you have a question about the ProcessMaster FEP630 and is the answer not in the manual?