F EM63 0 | E L EC TR OM AG N E T I C FLOWM E T E R | O I/F E M6 3 0-EN RE V. A
. . .7
Commissioning
. . .Hardware settings
Single-compartment housing
1
DIP switch, Write protection
Figure 61 Position of the DIP switch
The DIP switches are used to configure specific hardware
functions. The power supply to the transmitter must be
briefly interrupted or the device reset in order for the
modified setting to take effect.
Write-protect switch
When write protection is activated, device parameterization
cannot be changed via the LCD indicator. Activating and
sealing the write protection switch protects the device
against tampering.
Number
On
Off
Function
Write protection active
Write protection deactivated.
Configuration of digital outputs V1 / V2 or V3 / V4
1
NAMUR rotary switch
Figure 62 Position of rotary switch on the plug-in card
The configuration (NAMUR, optoelectronic coupler) for the
digital output on the plug-in card is set via a rotary switch on
the plug-in card.
Number
On
Off
Digital output V1 / V2 or V3 / V4 as NAMUR
Digital output V1 / V2 or V3 / V4 as
optoelectronic coupler output.
Function
output.