ABB ProcessMaster FEM630 Operating Instructions Manual page 85

Electromagnetic flowmeter
Hide thumbs Also See for ProcessMaster FEM630:
Table of Contents

Advertisement

FE M63 0 | E L EC TR OM AG N E T I C FLOWM E T E R | O I/F E M6 3 0-EN RE V. A
8
. . .Operation
. . .Parameter descriptions
Menu / parameter
Device Setup / . . . T ransmitter / . . .Feature Settings
Backwards Comp.
Backwards Comp. Code
Verifying Capab.
Verifying Capab.Code
Batchmode
Batchmode Code
Expert Diagn.
Expert Diagn. Code
SIL Comp.
SIL Comp. Code
Device Setup / ...System Zero
Manual Adjust
Auto Adjust
Description
Indicator as to whether the backward compatibility function has been activated.
Set the device-specific code for activating the function. To use this function subsequently, contact the ABB
service team or sales organization.
After entering the code, restart the device (e.g. using the parameter 'Device Reset' or by briefly switching off
the power supply).
Indicator as to whether the verification function has been activated.
Set the device-specific code for activating the verification function. To use this function subsequently, contact
the ABB service team or sales organization.
After entering the code, restart the device (e.g. using the parameter 'Device Reset' or by briefly switching off
the power supply).
Indicator as to whether the filling function has been activated.
Set the device-specific code to activate the filling function. To use this function subsequently, contact the ABB
service team or sales organization.
After entering the code, restart the device (e.g. using the parameter 'Device Reset' or by briefly switching off
the power supply).
Indicator as to whether advanced diagnosis functions such as gas bubble or conductivity have been activated.
Set the device-specific code for activating the advanced diagnosis function. To use this function subsequently,
contact the ABB service team or sales organization.
After entering the code, restart the device (e.g. using the parameter 'Device Reset' or by briefly switching off
the power supply).
Indicator as to whether the SIL function is active.
Set the device-specific code for activating the SIL function. To use this function subsequently, contact the ABB
service team or sales organization.
After entering the code, restart the device (e.g. using the parameter 'Device Reset' or by briefly switching off
the power supply).
Sets the value for zero point adjustment in % of Q
Manual adjustment: -2 to +2 inch/s (−50 to +50 mm/s)
Starts the automatic zero point balancing using
Note
Prior to starting the zero point adjustment, make sure that:
• There is no flow through the sensor (close all valves, shut-off devices etc.)
• The sensor must be filled completely with measuring medium for measurement.
DN
max
. Automatic zero point balancing takes approx. 60 seconds.