8
. . .Operation
. . .Parameter descriptions
Menu / parameter
Diagnostics / . . .Diagnosis Values
All values in this menu are for informational and service purposes only.
Inhouse Temperature
Inhouse Temp MaxPeak
Inhouse Temp MinPeak
Driver Current
Signal Region on ADC
Coil DAC Preset
DC Feedback DAC
ADC Errors
Device Restart Count
Diagnostics / . . . . . .Simulation Mode
Simulation Switch
Off
Qm Massflow [unit]
Qm Massflow [%]
Qv Volumeflow [unit]
Qv Volumeflow [%]
Conductivity[µS/cm]
Curr.Out 31/32
Curr.Out V1/V2
Curr.Out V3/V4
Dig.Out 41/42 State
Dig.Out 41/42 Freq.
Dig.Out 41/42 Pulse
Dig.Out 51/52 State
Dig.Out 51/52 Pulse
Dig.Out V1/V2 State
Dig.Out V3/V4 State
Dig.In V1/V2 State
Dig.In V3/V4 State
Hart Frequency
Description
Display of temperature value within the transmitter housing.
Display of the drive current of the sensor coil.
Display of the measuring signal within the A / D converter input.
(-100 % to +100 %)
Display of the D / A converter for coil drive.
D / A converter feedback value.
A / D converter error
Number of device restarts (boots).
Manual simulation of measured values. After selecting the value to be simulated, a corresponding parameter is
displayed in the menu 'Diagnostics / ......Simulation Mode'. The simulation value can be set here.
The output values correspond to the simulated flowrate entered.
The 'Configuration' information is displayed in the lower line of the display.
Only one measured value / output can be selected for simulation.
After power-up / restart of the device, the simulation is switched off.