. . .7
Commissioning
...Parameterization via the menu function
Easy Setup
Easy Setup
Pulse Width
Next
22 Use
to call up the edit mode.
23 Use
/
to select the desired pulse width for the
pulse output..
24 Confirm the selection with
Easy Setup
Damping
25.000 ugal/s
Next
25 Use
to call up the edit mode.
26 Use
/
to set the desired damping.
27 Confirm the selection with
Easy Setup
Dig.Out 41/42 Mode
Next
28 Use
to call up the edit mode.
29 Use
/
to select the desired operating modeOff,
Logic, Pulse, Frequency for the digital output.
30 Confirm the selection with
Easy Setup
Curr.Out at Alarm
Next
31 Use to call up the edit mode.
32 Use
/
to select the desired alarm mode.
33 Confirm the selection with
30.00 ms
Edit
.
Edit
.
Pulse
Edit
.
High Alarm
Edit
.
Easy Setup
Low Alarm
Next
34 Use
to call up the edit mode.
35 Use
/
to set the desired current for Low Alarm.
36 Confirm the selection with
Easy Setup
High Alarm
Next
37 Use
to call up the edit mode.
38 Use
/
to set the desired current for High Alarm.
39 Confirm the selection with