. . .Diagnosis / error messages

Error messages

The error messages are divided into four groups in accordance with the NAMUR classification scheme.
Errors
Error no . / Range
Text on the LCD display
ADC overrange alarm.
F099.042 / electronics
Noise too high. Check applicati.
Call service.
No Frontend Board detected.
F098.011 / electronics
Wrong connection. Defect
Frontend. Check wiring.
Coil regulation error.
F097.029 / Electronics
Check wiring of sensor coils. Call
service
SIL self check alarm.
F096.043 / Electronics
Call service.
Coil isolation alarm.
F095.036 / Electronics
Call service.
Safety Alarm Curr. Out 31 / 32
F094.021 / electronics
SIL function detects error. Call
Service.
Electrode short cuircit.
F093.032 / Electronics
Check wiring of sensor electrode.
Call service.
Electrode open cuircit.
F092.033 / Electronics
Check wiring of sensor electrode.
Call service.
Coil wiring error.
F091.030 / Electronics
Check wiring of sensor coils. Call
service.
ADC RX210 com. error.
F090.035 / Electronics
Call service.
Cause
Electrode signal overranges max. ADC limits.
No flow measurement possible.
Frontend board or motherboard hardware
defective.
Wrong or no connection between frontend
board and motherboard.
Incorrect coil wiring (M1 / M2 terminals) or
cable break / short-circuit or defective coil
fuse or moisture in terminal compartment.
The SIL monitoring function has detected a
transmitter error.
Defective coil or incorrect wiring (short-
circuit between M1 / M2 and GND).
Flooded sensor
The μController of the current output has
detected relevant SIL errors.
Wrong wiring or electrode leakage or short-
circuit of the electrode signal line and shield
or flowmeter flooded.
Wrong electrode wiring or break in electrode
signal line.
Incorrect coil wiring (M1 / M2 terminals) or
cable break / short circuit or defective coil
fuse or moisture in terminal compartment.
Bad EMC environment or defective
component.
Remedy
If the tube is empty, make sure the empty
tube detection is activated.
Make sure that the actual flow rate does not
up-scaled the configured flow rate.
Contact the service department
Check the wiring in the terminal
compartment and in the transmitter housing.
Contact the service department.
Check for incorrect coil wiring (M1 / M2
terminals) or cable break / short-circuit or
defective coil fuse or humidity in the terminal
compartment.
Contact the service department
Contact the service department
Contact the service department
Check for incorrect wiring or electrode
leakage or short-circuit of the electrode
signal line and the shield or flooded sensor.
Check for incorrect electrode wiring or break
in electrode signal line.
Check for incorrect coil wiring (M1 / M2
terminals) or cable break / short-circuit or
defective coil fuse or humidity in the terminal
compartment.
Replace the electronics unit or contact ABB
Service.