7
Nonlinear error compensation (NLEC)
For Nonlinear error compensation (NLEC) the measuring range is divided into a
grid by defining up to 99 supporting points. For each of the identically-sized surface
segments a specific compensation factor is determined by comparing nominal and
actual values (the measured values).
The following options are available for measuring the nominal and actual values of
the supporting points:
Create the table of supporting points manually and enter actual values
Create the table of supporting points manually and determine actual values in a
teach sequence
Read deviations of the calibration standard and enter actual values, determine
these in a teach sequence or import the values (.txt)
Import formats:
.txt format: contains actual values of the supporting points
Extended .txt format: contains deviations of the theoretical nominal value
Import a table of supporting points with nominal and actual values (.xml)
Import format
.xml format: corresponds to the standard export format of the product
94
If you read the deviations of a calibration standard or import a table of
supporting points and the number or distances of the supporting points
are different from the existing table of supporting points, a new table of
supporting points is generated. The original table of supporting points is
overwritten.
Tap Settings in the main menu
Open in the sequence
Axes
General settings
Error compensation
Nonlinear error compensation (NLEC)
Deactivate Compensation with the ON/OFF sliding switch
Enter the Number of supporting points for each axis
Enter the Spacing of the supporting points for each axis
To import existing calibration data, tap Read deviations of
calibration standard
Further information:
format", Page 95
Navigate to the desired folder
Tap the desired file (.acf)
Tap Select
The calibration data of the calibration standard are read
To import a table of supporting points, tap Import table of
supporting points
Further information:
points in .txt format", Page 96
Navigate to the desired folder
Tap the desired file (.xml or .txt)
Commissioning | Steps for commissioning
"Measuring calibration data in .acf
"Measuring the table of supporting
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 12/2018