11
Parameter
Error monitor
Counting direction
Settings for encoders with interfaces of the TTL type
Parameter
Encoder input
Incremental signal
180
Explanation
Monitoring of signal errors
Settings:
Off: Error monitoring not active
Amplitude: Error monitoring of signal amplitude
Frequency: Error monitoring of signal frequency
Frequency & amplitude: Error monitoring of signal amplitude
and signal frequency
Default value: Frequency & amplitude
A warning or error message is displayed if one of the
limit values for error monitoring is exceeded.
The limit values depend on the signal of the connected encoder:
Signal 1 Vpp, setting Amplitude
Warning with voltage ≤ 0.45 V
Error message with voltage ≤ 0.18 V or ≥ 1.34 V
Signal 1 Vpp, setting Frequency
Error message with frequency ≥ 400 kHz
Signal 11 µApp, setting Amplitude
Warning with current ≤ 5.76 µA
Error message with current ≤ 2.32 µA or ≥ 17 .27 µA
Signal 11 µApp, setting Frequency
Error message with frequency ≥ 150 kHz
Signal detection during axis movement
Settings:
Positive: The direction of traverse corresponds to the
counting direction of the encoder
Negative: The direction of traverse does not correspond to
the counting direction of the encoder
Default value: Positive
Explanation
Assignment of the encoder input to the axis of the product
Settings:
X21
X22
X23
"Product overview", Page 41
Further information:
Signal of the connected encoder
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 12/2018
Settings | Axes