8
Functions for measured value output
The following functions are available in the Inspector for measured value output:
Manual output of measured values: The user triggers transfer of the
measured values manually.
Touch-probe triggered output of measured values: The product automatically
transfers the measured values each time the stylus is deflected.
Continuous output of measured values: The product transfers the measured
values continuously at intervals of approx. 200 ms
The overview shows which contents can be transferred with the respective
functions:
Symbol
128
Function
Manual output of
measured values
Touch-probe
triggered output
of measured
values
Continuous output
of measured
values
You can use the Manual output of measured values and Touch-probe
triggered output of measured values in parallel. The Continuous
output of measured values cannot be combined with any of the other
measured value outputs.
See the
General Operation
the function elements for the measured value output.
Further information:
"Function elements", Page 68
As an alternative to the functions of the Inspector described here, you
can use the Trigger data transfer trigger function for the measured
value output.
Further information:
"Switching functions", Page 170
Current
position
Minimum
Maximum
chapter of this manual for a description of
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 12/2018