2
Installation of the Device

Site Ventilation

Site Ventilation
When CO
released through the pressure compensation opening into the operating room. 
Pressure compensation and any opening of the glass door/gas-tight screen during the operation of the
device will release very small quantities of CO
must be capable of carrying the released gas safely off into the open.
In addition, heat dissipating from the device during continuous operation may cause a change in the
room climate.
• Therefore, the
ventilation.
• Do not install the device in room recesses without ventilation.
• The room ventilation should be a technical ventilation that complies with the national
requirements for laboratories or some other suitable ventilation system with appropriate capacity.

Space Requirements

WARNING EMERGENCY STOP!
The mains socket must be accessible at all times for the case of an EMERGENCY STOP. The
mains socket must be easily identifiable by the user and freely accessible at all times.

Figure 2-1. Device Dimensions

Note Accessibility of the devices: 
To ensure the accessibility for care and maintenance works, keep larger side and rear distances. The
swivel-out for units with sixfold gas tight screen with Cell Locker must have a clearance to the wall
directly at the hinge side of >= 31 cm / 12.2".
Heracell Vios 160i / 250i CR
2-2
/O
/N
is supplied, the work space of the incubator is slightly pressurized. The pressure is
2
2
2
Heracell Vios 160i / 250i CR
/O
/N
into the operating room. The room ventilation
2
2
2
must only be installed in rooms with sufficient
All dimensions in mm
Thermo Scientific