Gettrighndlr - Racal Instruments 1260-00C User Manual

Slot 0 / resource manager
Table of Contents

Advertisement

GetTrigHndlr

Local Command Set 4-50
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
and the value of <protocol> specifies the protocol
to use:
Value
2
3
4
5
6
Response:
Program response: 0
Console response:
Trigger sense enabling (line = <line
text>, protocol = <protocol text>)
complete.<CRLF>
where the meaning of <line text> corresponds
to the value of <line> as follows:
Value of <line>
0 to 7
8 to 9
50
60
and the meaning of <protocol text>
corresponds to the value of <protocol> as
follows:
Value of <protocol>
2
3
4
5
6
Example:
Enable sensing of TTL line 4 for SEMI-SYNC
protocol.
EnaTrigSense 4, 5
Purpose:
Get the address of the current TTL/ECL trigger,
counter, or tick timer interrupt handler for a
specified trigger source.
Query
Syntax:
GetTrigHndlr <line>
1260-00C User Manual
Protocol
START
STOP
SYNC
SEMI-SYNC
ASYNC
Value of <line text>
TTL <line>
ECL (<line> - 8)
TCNTR
TICK 1
Value of <protocol text>
START
STOP
SYNC
SEMI-SYNC
ASYNC

Advertisement

Table of Contents
loading

This manual is also suitable for:

1260-00b

Table of Contents