Switch S15
Switch S59
OFF
ON
OFF
ON
Slot 0 Resource Manager
Operation
Front Panel LED
Indications for RM
Operation
Table 3-8. Front Panel LED Indications For RM Operation
Sequence
FAILED
1
OFF
2
ON
3
ON
4
OFF
5
OFF
ON
ON
OFF
Operation 3-14
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 3-8, CLK10 Routing Options
OFF
CLK10 sourced from onboard clock.
OFF
CLK10 and EXT CLK connector sourced from onboard clock.
ON
CLK10 sourced from an external clock via the EXT CLK
connector.
ON
Invalid. Do not use this setting.
At start-up, a 1260-00C configured as a Slot 0 Resource Manager
performs its self-tests, executes the RM functions, and enters its
normal mode of operation.
The five front panel LEDs are SYSFAIL, FAILED, TEST, ON LINE,
and ACCESS. The 1260-00C uses the FAILED, TEST, and ON
LINE LEDs to indicate the progress of its self-initialization,
self-test, and RM functions. The LED indications are shown in
Table 3-9. A successful system start-up will sequence through the
first five states. The point of failure is indicated for states in which
the FAILED LED is lit for an extended period of time. The LED
indications are identical for Slot 0/Non-Slot 0 Resource Manager
operation.
TEST
ON LINE
OFF
OFF
OFF
OFF
ON
OFF
ON
ON
OFF
ON
ON
ON
OFF
ON
ON
OFF
1260-00C User Manual
Function
State
No power
Failed before self-
test
In self-initialization
Failed in self-test
In self-test
Performing RM
Online
Failed
Failed while in RM
Failed
Failed while online
In non-volatile
configuration, or
diagnostics mode
Point of Failure