Racal Instruments 1260-00C User Manual page 125

Slot 0 / resource manager
Table of Contents

Advertisement

Local Command Set 4-76
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Syntax:
DCISetup? <logical address>,
<Commander's logical address>, <start
block>, <number of blocks>, <stack
size>, [, <Servant1>, [<Servant2>,...,
<ServantN>]]
The DCISetup? query provides the 1260-00C with
the information it needs to prepare for executing a
DCIDownLoad command. This command is
performed separately from the DCIDownLoad
command so that the download parameters can be
validated before the object code download is
initiated.
The 1260-00C interprets the DCISetup? query
parameters as follows:
The DCI is to be assigned Logical Address
<logical address>, and granted to the
Commander at <Commander's logical
address> as a Servant.
Up to <number of blocks> of DCI code and
data are to be loaded into CI RAM area starting
at <start block>.
A stack of size <stack size> words is to be
allocated for the CI worker process. If <stack
size> is less than 1024, a stack size of 1024
words (2 kilobytes) is to be allocated for the CI.
<Servant1> through <ServantN> are to be
granted to the CI as Servants.
Any GPIB address links to <Servant1>,
<Servant2> through <ServantN> must be
disconnected before the DCI is downloaded. You
can delete the links by using the SaDisCon or
LaSaddr commands.
Response: Program response:
0 <CRLF>
0 is returned to report the successful completion of
the DCISetup? command. Any errors are
reported in the form of an error code.
Console response:
DCI parameters OK, Ready to
1260-00C User Manual