Racal Instruments 1260-00C User Manual page 106

Slot 0 / resource manager
Table of Contents

Advertisement

1260-00C User Manual
TrigExtConf
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
input to counter:
Value
0 to 7
8 to 9
70
71
and <count> specifies the number of input pulses
to count before terminating.
Response:
Program response: 0
Console response:
CNTR configured (mode = <mode text>,
source = <source text>, count =
<count>).<CRLF>
where the meaning of <mode text> corresponds
to the value of <mode>:
Value of <mode>
0
2
3
and the meaning of <source text> corresponds
to the value of <source>:
Value of <source> Value of <source text>
0 to 7
8 to 9
70
71
Example :
Configure the counter count 25 assertions on TTL
trigger line 5 (the <protocol> parameter when
calling EnaTrigSense will determine whether the
counter accepts SYNC or SEMI-SYNC assertions).
TrigCntrConf 0, 5, 25
Purpose:
Configure the external trigger (GPIO) lines.
Query
Syntax:
TrigExtConf <extline>, <mode>
where the value of <extline> is the trigger line to
Trigger Line
TTL trigger lines 0 to 7
ECL trigger lines 0 to 1
CLK10
EXTCLK
Value of <mode text>
INIT
RELOAD
DISABLE
TTL <line>
ECL (<line> - 8)
CLK10
EXTCLK
Local Command Set 4-57