A16?

A24

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Command
Syntax:
A16
<A16 address>, <word value>
Action: Write <word value> to <A16 address>.
Example:
Write A502h to VXI A16 address 4305h.
A16 #h4305, #hA502
Purpose:
Read word value from VXI A16 address space.
Query
Syntax:
A16?
<A16 address>
Response:
Program response:
<word value><CRLF>
Console response:
Value <hex word value> (<word value>)
read from A16 address <A16 hex
address> (<A16 address>)<CRLF>
where <word value> is in decimal integer format,
<hex word value> is in C language
hexadecimal format, <A16 hex address> is in C
language hexadecimal format, and <A16
address> is in decimal integer format.
Example:
Read the ID register of Logical Address 16.
A16? #hc400
Purpose:
Write a 16-bit value into VXI A24 space.
Command
Syntax:
A24
<A24 address>, <word value>
Note that the <A24 address> has a valid range
of 2097152 to 14680062 (#h200000 to
#hE7FFFE).
Action:
Write <word value> to <A24 address>.
Example:
Write the value A502h to VXI A24 address
504305h.
Local Command Set 4-45