Wsresp; Wsstr - Racal Instruments 1260-00C User Manual

Slot 0 / resource manager
Table of Contents

Advertisement

1260-00C User Manual

Wsresp?

WSstr

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Purpose:
Read a 16-bit Word Serial response to a previously
sent query.
Query
Syntax:
WSresp?
Action:
Reads and returns the response of the device at
<log addr>.
Response:
Program response:
<hex value><CRLF>
where <hex value> is the hexadecimal value of
the Data Low register response.
Console response:
Logical Address <log addr> returned
response 0x<hex value><CRLF>
Example:
Read the 16-bit response to a previously sent Word
Serial query from Logical Address 3.
WSresp? 3
Purpose:
Send a device-dependent command string to a
Message-Based device.
Command
Syntax:
WSstr
where <string> is an ASCII character sequence
enclosed by double quotation marks (").
The following sequences of characters within the
<string> parameter are special cases and will be
interpreted as follows:
\n
\r
\\
\xHH any binary 8-bit value where HH is
<log addr>
<log addr>, <string>
linefeed (LF)
carriage return (CR)
backslash (\)
the ASCII hexadecimal
representation of that value
Local Command Set 4-67