Lasaddr - Racal Instruments 1260-00C User Manual

Slot 0 / resource manager
Table of Contents

Advertisement

1260-00C User Manual

LaSaddr

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
These commands and queries configure and report the
relationships between VXI logical addresses and GPIB addresses.
Determine the 1260-00C's primary address when used for multiple
GPIB secondary addressing by using the Primary? query from
the serial port. Determine the relationships between GPIB
addresses and VXI logical addresses by using the Saddrs?
query followed by SaddrLa? queries, or by using the RM
information query Laddrs? followed by LaSaddr? queries.
Assign GPIB address links to Message-Based Servants of the
1260-00C with the LaSaddr command. The SaDisCon
command deletes all GPIB address links except the link to the
1260-00C local commands.
The letters SA or SADDR in this chapter originally stood
for GPIB Secondary Address.
configured to handle multiple primary addresses as well.
The terminology has been left the same to maintain
backwards compatibility.
Purpose:
Attach or detach a GPIB address to a logical
address.
Command
Syntax:
LaSaddr
address>
Action:
If <GPIB address> is not equal to 255, attach
<GPIB address> to <logical address>.
If <GPIB address> is equal to 255, release
<GPIB address> currently attached to
<logical address>.
Attaching a GPIB address to a logical address that
already has a GPIB address will cause the first
GPIB address to be replaced by the new GPIB
address.
Attempting to release or change a GPIB address
will result in a Delete I/O Link error if any of the
following conditions are true:
The GPIB address does not exist.
NOTE
The 1260-00C can be
<logical address>, <GPIB
Local Command Set 4-33

Advertisement

Table of Contents
loading

This manual is also suitable for:

1260-00b

Table of Contents