Figure 3-7, Address Modifier Signals Switch Settings

1260-00C Start-Up
Mode Configuration
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Start-up mode switches S19 and S20 control the 1260-00C
operation mode at system start-up. They select one of four
modes, as shown in Figure 3-8. The four possible modes of start-
up are:
1. 488-VXI Runtime System Mode - The start-up mode for normal
operation in a VXI system, and is configured at the factory to
start up in this method. The remainder of this section
contains a description of this operation.
2. Non-Volatile Configuration Mode - Edits the contents of the
non-volatile configuration parameter memory. See Chapter
5, Non-Volatile Configuration for more information.
3. Diagnostics Mode - Performs extensive offline diagnostic tests
on the 1260-00C. See Chapter 6, Diagnostic Tests, for a
description of the self-tests.
Operation 3-11