DCON?

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
DCON? query returns configuration information about a device.
The DRES? query is used to perform a soft-reset sequence on a
device.
The RREG? query and WREG command are used to peek (read
from) and poke (write to) registers on a VXI device.
Purpose:
Return system configuration information about a
device or all devices.
Query
Syntax:
DCON? [<logical address>]
(If <logical address> is omitted, DCON?
returns the configuration information for all devices.)
Response:
Program response:
<la1>,<cla>,<IHANS>,<INTS>,<status>,<s
state>,<com><CRLF>
Console response:
Device configuration at Logical
Address <la>:<CRLF>
<CRLF>
Commander's Logical Address:<cla><CRLF>
Interrupt Handlers
Interrupters
Passed/Failed/Ready
Device Substate
Manufacturer Specific Comment
The mnemonics have the following meanings:
la
cla
IHANS
INTS
:<IHANS><CRLF>
:<INTS><CRLF>
:<status><CRLF>
:<sstate><CRLF>
Device's logical address
Commander's logical address
Interrupt handler levels used by this
device where IHANS is a 7-digit
binary representing the seven VXI
interrupt levels and a 1 in each
position, meaning Interrupt Handler
present
Interrupter levels used by this device
Local Command Set 4-27
:<com><CRLF>