Dres; Rreg - Racal Instruments 1260-00C User Manual

Slot 0 / resource manager
Table of Contents

Advertisement

1260-00C User Manual

DRES?

RREG?

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Purpose:
Perform a soft-reset sequence on a device.
Query
Syntax:
DRES? <logical address> [, <sysfail
flag>]
If the device stays failed for five seconds after the
soft-reset sequence, <sysfail flag> determines whether or
not the device is kept sysfail-inhibited.
Response:
Program response:
<status>
Console response:
Logical Address <logical address> is
<status>.
<state>.<CRLF>
where <status> is one of the following:
PASS
FAIL
IFAIL
READY
and <state> is one of the following:
ON
OFF
Example:
Soft-reset device at Logical Address 3.
DRES? 3
Purpose:
Read a 16-bit VXI register from a device.
Query
Syntax:
RREG? <logical address>, <reg offset>
where <logical address> is the device to read
from and <reg offset> is the number of bytes to
offset from the base of the VXI registers for that
device.
Response:
Program response:
NOTE
<CRLF>
SYSFAIL Inhibit is
Local Command Set 4-31