Table 3-15, Clk10 Routing Options - Racal Instruments 1260-00C User Manual

Slot 0 / resource manager
Table of Contents

Advertisement

1260-00C User Manual
Jumper/Switch
Switches S9 and S15
Switch S22
Switch S23
Switch S24
Logical Address
Slot 0 Model Code
Servant Area Size
Switch S14
Switch S17
OFF
OFF
ON
ON
Slot 0 Message-Based
Device Operation
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Position
See Table 3-15.
ON
ON
ON
See Chapter 5.
See Chapter 5.
See Chapter 5.

Table 3-15, CLK10 Routing Options

OFF
CLK10 sourced from onboard clock.
ON
CLK10 and EXT CLK connector sourced from
onboard clock.
OFF
CLK10 sourced from an external clock via the
EXT CLK connector.
ON
Invalid. Do not use this setting.
At start-up, a 1260-00C configured as a Slot 0 Message-Based
device performs its self-tests. It then waits until it receives its
Device Grant (if any) and Begin Normal Operation Word Serial
commands. The RM grants any logical addresses to the
1260-00C that reside within its Servant area. When the 1260-00C
responds to the Begin Normal Operation command, it enters the
normal mode of operation.
After the 1260-00C Passed bit is set, the RM can manipulate or
Function
CLK10 routing options.
VXI BTO enabled.
Bus arbiter and SYSCLK
enabled. CLK10 sourcing
backplane.
MODID pulled up
Logical address is not equal
to 0. Set in non-volatile
configuration.
Model code is set to the Slot 0
value. Set in non-volatile
configuration.
Set appropriate Servant area
size. Set in non-volatile
configuration .
Function
Operation 3-23