Slot 0 / resource manager
Table of Contents

Advertisement

TrigToREQT

Local Command Set 4-60
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Value of <mode>
0
1
2
3
and the meaning of <source text> corresponds
to the value of <source>:
Value of <source> Value of <source text>
40 to 49
70
71
Example 1 :
every 6.55 milliseconds by dividing down CLK10 as
an input. Call EnaTrigSense to start the tick
timers and enable interrupts.
TrigTickConf 0, 70, 16, 0
Example 2 : Configure the tick timers to output a continuous
9.765-kHz square wave on TICK1 output and a 1.25
MHz clock on TICK2 output by dividing down
CLK10 as an input. Call SrcTrig to start the tick
timers.
TrigTickConf 0, 70, 10, 3
Purpose:
Map trigger interrupt to GPIB SRQ condition (REQT
generation for a particular GPIB address).
Command
Syntax:
TrigToREQT <la>, <line>
where <la> identifies the device for which to
assert SRQ, and <line> is the trigger line for
which to map the interrupt, where the value of
<line> corresponds to the trigger line or
counter/tick:
Value
0 to 7
8 to 9
50
60
Action:
The 1260-00C is set up to assert SRQ for a device
attached to a GPIB address for a given trigger line's
1260-00C User Manual
Value of <mode text>
INIT w/ROLL
INIT w/NOROLL
RELOAD
DISABLE
GPIO (<source> - 40)
CLK10
EXTCLK
Configure the tick timers to interrupt
Trigger Line
TTL trigger lines 0 to 7
ECL trigger lines 0 to 1
TIC counter
TIC TICK1 tick timer