11.4

Replacing a terminal block socket module

Note
This section only applies to sensors with Kx approval codes.
The terminal block socket module is shown in
module, remove the junction box from the sensor adapter.
Figure 11-2: Socket Module—Terminal Block
A. Mounting screws:
11.4.1
Removing a terminal block socket module
1. Disconnect power to the transmitter and the remote cabling connected to the
2. Remove the junction box cover to gain access to the remote cabling.
3. To disconnect the terminal block from the junction box housing, remove the two
4. Pull up on the terminal block to expose the socket module base.
5. To remove the socket module, loosen the two mounting screws and pull up on the
6. When removing the socket module, do not pull on the wires.
11.4.2
Installing a terminal block socket module
1. Insert the new terminal block socket module, press the base into its keyed position,
Reference Manual
2X—standard
4X—with I.S. divider
terminal block.
mounting screws and the two divider mounting screws (if applicable).
socket module from the base.
and tighten the two mounting screws.
Figure
11-2. To gain access to the socket
A
Maintenance
November 2019
117