6 Troubleshooting

6.1 General Notes

The AFC-3000 is both easy to operate and reliable. When the fraction collector does not
function properly, isolate the problem to determine if it originates in the Chromeleon
computer, the analytical instrument, the communications cable, or the fraction collector
itself. The following sections help you to identify and eliminate the source of problems that
may occur during the operation of the AFC-3000.
If you determine the problem is in the AFC-3000, check the power system, the
communications interface, or the diverter valve assembly to locate the cause of the problem
and resolve it.
If a problem occurs, always check the Chromeleon audit trail for messages (→ section 6.3).
If you cannot solve a problem using the steps given in this chapter, please contact Thermo
Fisher Scientific Service for Dionex HPLC Products.

6.2 Power and Communication Problems

The following table provides information about operating problems and lists probable
causes, as well as remedial actions.
Problem
The blue LED
power indicator is
off.
The AFC-3000 does
not work correctly
when controlled by
Chromeleon.
Operating Instructions
Probable Cause
The instrument is not connected
to the mains.
The power is turned off.
The power switch of the external
power supply is turned off.
An error occurred in the
electronic system.
There is no connection between
the fraction collector and the
Chromeleon computer.
UltiMate 3000 Series:
AFC-3000 Automated Fraction Collector
Remedial Action
Connect the power cord
(→ section 3.4.1; page 35).
Turn on the fraction collector power.
Turn on the power switch of the
external power supply (if equipped).
Contact Service.
Check the USB or RS-232 cable and
connection to the computer. If the cable
is connected, ensure that it is tightened
properly.
Verify that the cable is connected to
COM 1 port on the AFC-3000 rear
panel, or to the appropriate module
USB hub if connecting in that manner
(→ section 3.4; page 35).
Test the interface (→ section 4.3;
page 43).