9. Install the diverter valve connections (→ section 3.3.2, page 28). A low-flow inlet
capillary is shipped with the kit.
10. Verify that the alignment is still correct (→ page 44).
Tips: After installing the low-flow inlet capillary, always determine the delay time or
volume. Note that the carry-over volume of the kit for low flow rates is smaller
than with the standard configuration.
Exclusively use the low-flow drop former within the specified flow rates (up to
10 mL/min). At higher flow rates, the jet from the drop former may cause
sprays or spillages.

7.6 Replacing the Diverter Valve

Description
Diverter valve with 1.0 mm ID PEEK drop former
(up to 150 mL/min)
Diverter valve with 0.4 mm ID drop former (up to10 mL/min),
including tools and tubing
The rear cable is pre-installed on the diverter valve.
1. Shut down and unplug the fraction collector. Set the pump flow to 0.
2. Carefully move the diverter valve assembly along the Y arm to the front.
3. Remove the sample inlet and waste tubing and unscrew the current drop former from
the diverter valve.
4. Loosen the two diverter valve screws by using the hexagon wrench shipped with the
kit (→ Fig. 28).
Hexagon
wrench
Diverter valve
screws
Operating Instructions
Fig. 28: Dismounting the diverter valve
UltiMate 3000 Series:
AFC-3000 Automated Fraction Collector
Part No.
6702.0500
6702.0300
Page 67