Tip: To prevent the universal drip tray from accidentally being lifted out of its
proper position, always remove the rack adapter toward the side as shown here:

3.3.6 Adjusting the Drop Former Height

1. Grasp the diverter valve assembly firmly by the vertical stainless steel element while
rotating the adjustment locking lever upward. This will unlock the slider assembly,
allowing you to move it up and down freely (→ Fig. 12).
2. Slide the assembly so that the drop former is several millimeters above the highest
point of the fraction tubes or vials. Adjust as needed.
3. Grasp the diverter valve assembly firmly by the vertical stainless steel element while
rotating the adjustment locking lever downward. This will lock the diverter valve
assembly so that its height is fixed. If the lever cannot be moved downward easily, the
diverter valve assembly may not be positioned properly in one of the grooves.
Operating Instructions
Fig. 11: Removing the rack adapter
UltiMate 3000 Series:
AFC-3000 Automated Fraction Collector
Page 33