3.3.1 Mounting the Diverter Valve Assembly

Mounting the diverter valve assembly on the fraction collector arm is the first major task.
The diverter valve assembly attaches to the fraction collector arm using two knurled
finger-tight screws.
1. Place the diverter valve assembly on the fraction collector arm and mount it with two
knurled finger-tight screws. No tools are required to fasten the screws. Fig. 3 shows
the assembly attached.
Knurled screws
2. Attach the mounting block to the AFC-3000 rear panel with the finger-tight mounting
block screws provided (→ Fig. 4).
3. Connect the diverter valve cable to the auxiliary I/O connection on the AFC-3000 rear
panel (→ Fig. 2, page 18). Tighten the knurled nut finger-tight.
4. Screw the PEEK drop former finger-tight into the diverter valve.
Operating Instructions
Fig. 3: Diverter valve assembly
Fig. 4: Attaching the mounting block
UltiMate 3000 Series:
AFC-3000 Automated Fraction Collector