5.2.4 Defining the Collect Mode

Define how the drop former travels across the racks.
Important:
Important:
1. Open the Commands dialog box (→ page 46).
2. Click the plus sign beside AFC.
3. Click CollectMode. In the CollectMode box, select the collect mode:
Vertical: The drop former follows a meander-shaped path. Be aware that this mode
minimizes tube change times, but negates imprinted tube numberings.
SawVertical (default): The drop former follows a saw-shaped path. The movement
matches the imprinted numbering on all rack types that are available from Thermo Fisher
Scientific.
Important: To avoid accidental contamination of empty tubes, never use
Important: Afin d'éviter contamination des tubes, n'utilisez pas le mode de
4. Click Execute.
Tip: Schott bottles will always be processed in SawVertical collect mode. The
collect mode setting is not supported with Schott bottles.
Page 50
The CollectMode must not be changed during fraction collection, or
during the course of a program file or sequence.
Ne changez pas le CollectMode au cours de la collection de fraction,
ou au cours d'un programme ou séquence.
Vertical
Fig. 19: Vertical and SawVertical collect modes
SawVertical collect mode in combination with Continue movement
mode.
collection SawVertical avec le mode de mouvement Continue.
SawVertical