Model 3418 GUSM
Table 5-6: Calibrate analog inputs and outputs
Daniel
Dat Points, options and guidelines
MeterLink
Display name
Current output
Data points affected:
Output setting
Options:
Guidelines
Current output
Data points affected:
Start (Stop) AO1
Options:
Guidelines
Current output
Data points affected:
Output setting
Options:
Guidelines
Note
The AO2 is only available with CPU module, part number
1-360-03-010.
AO1estModeOutputPercent
Enter the integer percentage of full scale for the analog output
within the range [0, 100 %] where 0% corresponds to the
minimum output (4 mA) and 100% corresponds to the
maximum output (20 mA).
The specified test value takes effect within one batch period of
clicking the Start button.
IsAO1EnableTest
Click on the StartAO1 button to enter the test mode (TRUE).
Note
the Start button will become the Stop button after clicking on it.
Click on the StopAO1 button to exit the test mode (FALSE).
Note
the Stop button will become the Start button after clicking on it.
The specified test value takes effect within one batch period of
clicking the Start button (see below).
AO2TestModeOutputPercent
Enter the integer percentage of full scale for the analog output
within the range [0, 100 %] where 0% corresponds to the
minimum output (4 mA) and 100% corresponds to the
maximum output (20 mA).
The specified test value takes effect within one batch period of
clicking the Start button.
Configurations
January 2019
53