If a log type to be read is configured so that unread records are not over-written, then
Daniel MeterLink queries the user as to whether or not that log's records should be marked
as 'read'.
Once the data collection is completed, the data is displayed in the Meter Archive Logs
dialog box one log type at a time. Select the log type to be displayed via the View log box.
The data may be sorted by selecting either Oldest first or Newest first in the Sort order
box.
6.1.7
Collecting logs
To collect Maintenance and Trend Logs for meter diagnostics
Procedure
1. Establish communication with the meter via Daniel MeterLink.
2. From Daniel MeterLink main window, select Logs/Reports > Maintenance Logs and
3. Set the log collection Duration (mins) (2 minutes is the default value which is
112
Reports. The Maintenance Logs and Reports dialog displays.
Figure 6-1: Maintenance Logs
normally adequate to capture a good snapshot of the meter's current diagnostic
parameters). To get a good representation of the meter performance, it is
recommended that there be at least 30 records (lines of data). The actual number
of records collected are a function of the type of communication (Serial or
Ethernet), Stack Size (if stacking is turned on), and the Collection speed selected.
Operations manual
DAN-20057316