Configurations
January 2019
3. The screen(s) displays the current (live) reading. When the input is stable, click the
4. Enter the correct value in the Actual edit box. Finally, the Finish screen shows the
5. Click Finish to write the values to the meter. If the meter is not configured to use
6. Enable the WRITE PROT. switch on the CPU Module to write-protect the
52
The Offset method requires a single calibration point and only affects the input's
calibration offset parameter.
The two-point calibration requires calibrating at two different points (ideally far
apart in value) and affects the input's calibration offset and gain parameters.
Depending upon the selected calibration method, either one or two screens will
follow when the Next button is clicked.
Hold button to freeze the current reading.
new calculated offset and gain values.
Figure 5-5: Offset calibration
live values from the analog input for calculation, a prompt displays asking if the
configuration should be changed to use the live input.
configuration.
Operations manual
DAN-20057316
Operations manual